TESA B&S S.A.
Rue du Bugnon 38
CH- 1020 Renens
Tél. (021) 634 15 51 Fax (021) 631 75 12
E-mail . [email protected]
Page : 39
INSTRUCTIONS DE REPARATION
REPARATURANLEITUNG
REPAIR INSTRUCTIONS
DOCUMENT No: 334.80051 V03
DATE :15/08/2002
15 .Software Mhcor
What is the Mhcor software used for?
With Mhcor you can :
1. Verify and/or correct the slope error (squareness) on X and Y.
2. Correct the internal and external diameter errors.
3. Linearize the column.
Who should use this software ?
All Brown & Sharpe SA approved engineers that follow-up the repair course M-H V10 or superior. This software can in no case be
either copied or handed over to a third person without the authorisation of B&S TESA SA.
What should happen with the SCS certificate delivered with the Micro-Hite after the utilisation of the Mhcor software. ?
This software modifies the correction parameters. For this reason, the certificate becomes invalid since it is only valid for a M-H
calibrated and certified in a SCS laboratory.
Responsibility of B&S TESA SA
B&S TESA SA will not be responsible for any damage due to incorrect use of the Mhcor software..
What material will be necessary ?
1. Certainly a laboratory with stable temperature of 20°C ±0.5
2. In order to correct the slope - a square and a certified granite support.
3. For the correction of diameter errors - a certified ETALON ring.
4.
In order to linearize the column - either a Koba step or certified steps covering 100% of the measuring range of the M-H.
5. PC 386 or superior with MSDOS
6. Cable RS-232 - B&S SA TESA No. 04761023
7. Software Mhcor V2.01 or superior
Installation of the MHcor software ?
It is not necessary to install the MHcor on the hard disk. It can be executed directly from the floppy disk.
If you absolutely wish to install it on the hard disk, you must create a directory and save the files in it.