background image

 

©2012 Terumo BCT, Inc.  

 

 

 

Page 12 of 14

 

 

 

 

 

PN 306620452

 

Replacing the Air Filter 

 

Step

 

Action

 

The air filter should be replaced on a regular basis. It should also be replaced if it is 
damaged or if excessive smoke is noticed during welding operation. 

Contact Terumo BCT Customer Support at 1-877-339-4228 to obtain replacement 
filters (SDOT0467). 

 

Turn off the power and unplug the power cord from the device.  

 

Facing the TSCD II device, remove the filter case from the lower right 
side of the device by pulling it gently to the right. 

 

Remove the air filter from the filter case gently, being careful not to bend 
the four claws that hold the air filter in place.   

 

Clean the filter case with a foam applicator, a cotton swab, or a cloth 
dampened with a mild detergent or isopropyl alcohol. Wipe dry.  

 

Place a new air filter in the filter case. Place the air filter flat on the square 
area and hold it in place using the four claws.  

 

Carefully insert the filter case into the right side of the device and push it 
in gently. Damage will occur if the filter case is not properly aligned or if 
you push too hard.  

 

 

Replacing the Fuses 

Step

 

Action

 

Turn off the power and unplug the power cord from the device. 

  

Locate the fuse holders at the back of the device above the handle. 

 

Using a flat-head screwdriver, turn the cover of the holder one-half turn 
counterclockwise. 

 

 

Содержание TSCD II

Страница 1: ...pport Shelves Maintenance Procedures Cleaning Surfaces Cleaning the Clamps Cleaning the Wafer Cassette Cartridge Compartment Replacing the Air Filter Replacing the Fuses Emptying and Cleaning the Wafe...

Страница 2: ...products Prepare an aliquot for pediatric use and divided units For example attach a syringe set or a set of bags to prepare aliquots of whole blood red blood cells or plasma Connect additional saline...

Страница 3: ...e inch beyond the outer edge of both clamps Close the clamp covers Press the START button to initiate the welding cycle After the device beeps and the message WELD COMPLETE OPEN CLAMPS appears on the...

Страница 4: ...ide the box forward to remove it from the device c Discard the used wafers into an appropriate container Caution Wafers that have been used on tubing that contained blood products should be considered...

Страница 5: ...the back edge of the new wafer cartridge cassette until it snaps into place 7 Press the RESET button to align the clamps and to advance an unused wafer if one is not already in place Note While the w...

Страница 6: ...inch beyond the outer edge of the clamps Performing a Weld Step Action 1 Press the START button The message Heating wafer shows on the LCD display After the wafer reaches the required temperature the...

Страница 7: ...lt in welds that seem misaligned This is generally acceptable as long as the weld opens easily and no leaks occur 3 Pinch or roll the tubing until the fluid pathway opens Caution Do not abuse the weld...

Страница 8: ...from the device Attaching the cover 1 Hold the cover in a vertical position over the back of device 2 Slide the cover down into the hinge sockets until the cover snaps into place Assembling and Attac...

Страница 9: ...d be pointing outward away from the center support of the shelves Attach the remaining three feet in the same way Attaching the bag support shelves to the device 1 Insert the rollers of the first bag...

Страница 10: ...es of the device Cleaning Surfaces Step Action Surface cleaning is recommended on an as needed basis Surfaces of the device may be cleaned with a cloth dampened with a neutral detergent or isopropyl a...

Страница 11: ...ps in the following areas The clamp covers The tubing holder slots and teeth Cleaning the Wafer Cassette Cartridge Compartment Step Action Wafer cassette cartridge compartment cleaning is recommended...

Страница 12: ...being careful not to bend the four claws that hold the air filter in place 4 Clean the filter case with a foam applicator a cotton swab or a cloth dampened with a mild detergent or isopropyl alcohol...

Страница 13: ...ay indicates DISPOSAL BOX FULL Normal operations cannot resume until the wafer disposal box has been emptied Removing and emptying the wafer disposal box 1 Pull on the insert on the front of the box t...

Страница 14: ...Note The TSCD II device has two photo sensors that align with the slots on the wafer disposal box when it is inserted properly The photo sensors send a message to the LCD display when the box needs to...

Отзывы: