background image

 

©2012 Terumo BCT, Inc.  

 

 

 

Page 8 of 14

 

 

 

 

 

PN 306620452

 

Removing and Attaching the Cover 

 

Step

 

Action

 

Removing the cover 

1 Open 

the 

cover. 

 

Gently pull the side walls of the cover away from the TSCD II device.  
The hinge pins on the cover are released from the hinge sockets on the 
device, and the cover can be removed. 

Note: 

When the device is in use, the cover should be either in the open 

position or removed from the device. 

 

Attaching the cover 

Hold the cover in a vertical position over the back of device. 

 

Slide the cover down into the hinge sockets until the cover snaps into 
place. 

 

 

Assembling and Attaching the Bag Support Shelves 

 

Note:  

Do not attempt to lift or transport the TSCD II device using the bag support shelves.  

 

Step

 

Action

 

Assembling the bag support shelves 

Gather the following items, which are included with the purchase of the 
TSCD II device: 

 

Two bag support shelves 

 

Two sets of feet 

 Four 

screws 

 

Lay the support shelves on a flat surface with the underside (center 
support) facing up.    

 

Содержание TSCD II

Страница 1: ...pport Shelves Maintenance Procedures Cleaning Surfaces Cleaning the Clamps Cleaning the Wafer Cassette Cartridge Compartment Replacing the Air Filter Replacing the Fuses Emptying and Cleaning the Wafe...

Страница 2: ...products Prepare an aliquot for pediatric use and divided units For example attach a syringe set or a set of bags to prepare aliquots of whole blood red blood cells or plasma Connect additional saline...

Страница 3: ...e inch beyond the outer edge of both clamps Close the clamp covers Press the START button to initiate the welding cycle After the device beeps and the message WELD COMPLETE OPEN CLAMPS appears on the...

Страница 4: ...ide the box forward to remove it from the device c Discard the used wafers into an appropriate container Caution Wafers that have been used on tubing that contained blood products should be considered...

Страница 5: ...the back edge of the new wafer cartridge cassette until it snaps into place 7 Press the RESET button to align the clamps and to advance an unused wafer if one is not already in place Note While the w...

Страница 6: ...inch beyond the outer edge of the clamps Performing a Weld Step Action 1 Press the START button The message Heating wafer shows on the LCD display After the wafer reaches the required temperature the...

Страница 7: ...lt in welds that seem misaligned This is generally acceptable as long as the weld opens easily and no leaks occur 3 Pinch or roll the tubing until the fluid pathway opens Caution Do not abuse the weld...

Страница 8: ...from the device Attaching the cover 1 Hold the cover in a vertical position over the back of device 2 Slide the cover down into the hinge sockets until the cover snaps into place Assembling and Attac...

Страница 9: ...d be pointing outward away from the center support of the shelves Attach the remaining three feet in the same way Attaching the bag support shelves to the device 1 Insert the rollers of the first bag...

Страница 10: ...es of the device Cleaning Surfaces Step Action Surface cleaning is recommended on an as needed basis Surfaces of the device may be cleaned with a cloth dampened with a neutral detergent or isopropyl a...

Страница 11: ...ps in the following areas The clamp covers The tubing holder slots and teeth Cleaning the Wafer Cassette Cartridge Compartment Step Action Wafer cassette cartridge compartment cleaning is recommended...

Страница 12: ...being careful not to bend the four claws that hold the air filter in place 4 Clean the filter case with a foam applicator a cotton swab or a cloth dampened with a mild detergent or isopropyl alcohol...

Страница 13: ...ay indicates DISPOSAL BOX FULL Normal operations cannot resume until the wafer disposal box has been emptied Removing and emptying the wafer disposal box 1 Pull on the insert on the front of the box t...

Страница 14: ...Note The TSCD II device has two photo sensors that align with the slots on the wafer disposal box when it is inserted properly The photo sensors send a message to the LCD display when the box needs to...

Отзывы: