©2012 Terumo BCT, Inc.
Page 11 of 14
PN 306620452
Cleaning the Clamps
Step
Action
Clamp cleaning is recommended on an as-needed basis.
1
Turn off the device and unplug the power cord.
2
Using either a foam applicator or cotton swab dampened with isopropyl
alcohol or an alcohol pad, wipe down the clamps in the following areas:
The clamp covers
The tubing holder slots and teeth
Cleaning the Wafer Cassette (Cartridge) Compartment
Step
Action
Wafer cassette (cartridge) compartment cleaning is recommended on an as-needed
basis.
1
Turn off the device and unplug the power cord.
2
Remove the wafer cassette (cartridge).
3
Using either a foam applicator or cotton swab dampened with isopropyl
alcohol or an alcohol pad, wipe down the compartment.
4
Ensure that all foreign matter is removed from the compartment.