background image

 

©2012 Terumo BCT, Inc.  

 

 

 

Page 7 of 14

 

 

 

 

 

PN 306620452

 

Inspecting the Integrity of the Weld  

 

Caution:

  Some residual blood may be encountered under normal operating conditions. Tubing that 

contains blood products should be handled appropriately, according to your institution’s SOPs. Residual 
blood may be located on waste segments, welded tubing, and clamps. 

 

Step

 

Action

 

Remove the welded tubing from the clamp slots. 

 

Inspect each weld carefully.  

Without opening the weld, rotate the tubing 360° and visually inspect it for 
alignment.  Each weld should be completely aligned. 

Caution:

 Misaligned welds may or may not be complete and should 

be thoroughly inspected for leaks. 

Note:

 Refer to Fig. 5-8 in the operating instructions for illustrations of 

acceptable welds.   

Note:

 Welding tubing of different diameters may result in welds that 

seem misaligned. This is generally acceptable, as long as the weld 
opens easily and no leaks occur.   

 

Pinch or roll the tubing until the fluid pathway opens. 

Caution:

 Do not abuse the weld by pulling or repeatedly bending it, 

since this could damage the weld. 

Note:

 If the weld is difficult to open, it could indicate either that the 

two sections of tubing that are welded together are incompatible or 
that there is an alignment problem.  

If the weld is misaligned and has not been opened, it may be possible to 
seal the tubing on both sides of the weld and prepare a new weld. 

 

Check each weld for leakage. 

Warning:

 If, for any reason, the weld leaks, the sterility of the product 

may be compromised. Follow your institution’s SOPs for evaluating 
and managing a potentially contaminated product and for 
decontaminating the work environment, if necessary. 

 

Carefully remove the waste segments and discard appropriately. Do not 
apply pressure or squeeze the segments, because the seals are 
temporary. 

 

 

Содержание TSCD II

Страница 1: ...pport Shelves Maintenance Procedures Cleaning Surfaces Cleaning the Clamps Cleaning the Wafer Cassette Cartridge Compartment Replacing the Air Filter Replacing the Fuses Emptying and Cleaning the Wafe...

Страница 2: ...products Prepare an aliquot for pediatric use and divided units For example attach a syringe set or a set of bags to prepare aliquots of whole blood red blood cells or plasma Connect additional saline...

Страница 3: ...e inch beyond the outer edge of both clamps Close the clamp covers Press the START button to initiate the welding cycle After the device beeps and the message WELD COMPLETE OPEN CLAMPS appears on the...

Страница 4: ...ide the box forward to remove it from the device c Discard the used wafers into an appropriate container Caution Wafers that have been used on tubing that contained blood products should be considered...

Страница 5: ...the back edge of the new wafer cartridge cassette until it snaps into place 7 Press the RESET button to align the clamps and to advance an unused wafer if one is not already in place Note While the w...

Страница 6: ...inch beyond the outer edge of the clamps Performing a Weld Step Action 1 Press the START button The message Heating wafer shows on the LCD display After the wafer reaches the required temperature the...

Страница 7: ...lt in welds that seem misaligned This is generally acceptable as long as the weld opens easily and no leaks occur 3 Pinch or roll the tubing until the fluid pathway opens Caution Do not abuse the weld...

Страница 8: ...from the device Attaching the cover 1 Hold the cover in a vertical position over the back of device 2 Slide the cover down into the hinge sockets until the cover snaps into place Assembling and Attac...

Страница 9: ...d be pointing outward away from the center support of the shelves Attach the remaining three feet in the same way Attaching the bag support shelves to the device 1 Insert the rollers of the first bag...

Страница 10: ...es of the device Cleaning Surfaces Step Action Surface cleaning is recommended on an as needed basis Surfaces of the device may be cleaned with a cloth dampened with a neutral detergent or isopropyl a...

Страница 11: ...ps in the following areas The clamp covers The tubing holder slots and teeth Cleaning the Wafer Cassette Cartridge Compartment Step Action Wafer cassette cartridge compartment cleaning is recommended...

Страница 12: ...being careful not to bend the four claws that hold the air filter in place 4 Clean the filter case with a foam applicator a cotton swab or a cloth dampened with a mild detergent or isopropyl alcohol...

Страница 13: ...ay indicates DISPOSAL BOX FULL Normal operations cannot resume until the wafer disposal box has been emptied Removing and emptying the wafer disposal box 1 Pull on the insert on the front of the box t...

Страница 14: ...Note The TSCD II device has two photo sensors that align with the slots on the wafer disposal box when it is inserted properly The photo sensors send a message to the LCD display when the box needs to...

Отзывы: