6
PRISM Field Upgrade Kit Instructions
7.
Locate the new Eye mezzanine board from the kit (part number 878-
1285-XX).
8.
Slide the Eye mezzanine board in at an angle from the side of the instrument
as shown in the following figure.
Figure 6: Eye mezzanine board placement
9.
Guide the BNC connectors through the back panel and line up the two board
connectors with the connectors underneath.
CAUTION.
Bending the Eye mezzanine board can damage the board or
instrument. Avoid fully inserting the back connector until the front connector is
properly aligned; this prevents the bending of the board.
10.
Begin by engaging the connector near the back panel as shown in the
following figure. Do not fully insert.
Figure 7: Board-to-board connector alignment