PRISM Field Upgrade Kit Instructions
3
Upgrade the PRISM MPI Eye mezzanine board
Complete the following steps to remove the old mezzanine board from the
instrument and install the new Eye mezzanine board:
1.
Set the instrument so that the bottom is down on a static-free work surface.
2.
Remove the top cover as follows:
a.
Use a TORX-driver with a T-10 TORX tip to remove the twelve screws
securing the cover to the instrument.
b.
Slide the cover back and then lift it from the back to remove it from the
instrument; set the cover aside.
Figure 1: Remove the cover