B-6
GASURVEYOR 7B-500R - USER HANDBOOK
Stainless Steel (Flue) Probe
This probe is also available as an accessory.
Fig. B-4 Flue Probe
When a flue sample is required, a Stainless Steel Open
Ended Probe (Part No. 13413) with Coalescing Filter
Assembly (Part No. 42215), incorporating Coalescing Filter
(Part No. 75174), is generally used. The purpose of the
filter assembly is to isolate any moisture which could affect
the sample reading. The flue probe and coalescing filter
assembly is generally used with the Probe Handle Assembly
(Part No. 12481), incorporating both Cotton and Hydrophobic
Filters, to prevent the ingress of dirt. The Sample Line (Part
No. 12712) connects the probe handle to the instrument
and allows the pump to draw the sample into the instrument
analysing chamber.