GASURVEYOR 7B-500R - USER HANDBOOK
ii
SAFETY
•
The instrument must be regularly serviced and calibrated by fully trained
personnel in a safe area.
•
The rechargeable battery pack must only be charged in a safe area.
•
The battery pack must be replaced in a safe area and fitted correctly
before use. Never use damaged battery packs or expose to extreme
heat. See Chapter 5: OPERATOR MAINTENANCE.
•
Only GMI replacement parts should be used.
•
If the instrument detects gas, follow your own organisation’s procedures
and operational guidelines.
•
The combustion chamber is a flameproof assembly and must not be
opened in the presence of a flammable atmosphere.
•
Gasurveyor 7B-500R instruments are certified as:
EEx iad IIC T4 -20
o
C<Tamb<50
o
C (-4
o
F<Tamb<122
o
F).
BAS01ATEX2292
II 2 G
UL 913 Class Groups A, B, C & D
•
This equipment is designed and manufactured to protect against
other hazards as defined in paragraph 1.2.7 of Annex II of the ATEX
Directive 94/9/EC.
Any right of claim relating to product liability or consequential damage to any third
party against GMI is removed if the warnings are not observed.
AREAS OF USE
Exposure to certain chemicals can result in a loss of sensitivity of the flammable
sensor. Where such environments are known or suspected it is recommended that
more frequent response checks are carried out. The chemical compounds that
can cause loss of sensitivity include Silicones, Lead, Halogens and Sulphur. Do
not use instrument in potentially hazardous atmospheres containing greater than
21% Oxygen. The enclosure material is polypropylene and must not be exposed to
environments which are liable to result in mechanical or thermal degradation or to
damage caused by contact with aggressive substances. Additional protection may
be required in environments where the instrument enclosure is liable to damage.
STORAGE, HANDLING AND TRANSIT
The batteries in the rechargeable pack contain considerable energy and care
should be taken in their handling and disposal. Battery packs should be removed
if the instrument is stored for longer than 3 months. The instrument is designed
to handle harsh environments. The sensing elements are sealed to IP54 and the
rest of the instrument to IP64. If not subject to misuse or malicious damage, the
instrument will provide many years of reliable service. The instrument contains
electrochemical sensors with a life of 2 years. Under conditions of prolonged storage
the sensors should be removed. The sensor contains potentially corrosive liquid
and care should be taken when handling or disposing of the sensor, particularly
when a leak is suspected.
Содержание GASURVEYOR 7B-500R
Страница 1: ...Gasurveyor 7B 500R User Handbook ...
Страница 10: ...viii GASURVEYOR 7B 500R USER HANDBOOK ...
Страница 22: ...2 10 GASURVEYOR 7B 500R USER HANDBOOK ...
Страница 34: ...3 12 GASURVEYOR 7B 500R USER HANDBOOK ...
Страница 50: ...8 2 GASURVEYOR 7B 500R USER HANDBOOK ...
Страница 62: ...xii GASURVEYOR 7B 500R USER HANDBOOK ...
Страница 63: ......