Maintenance
Model 6750
Teledyne Analytical Instruments
54
5.3.4 D. C. Power Supply
The power supply has no field adjustment. If malfunctioning,
replace the module (P/N ST15083) using the directions below. See
Figure 5-6.
Figure 5-6: DC Power Supply Module
To replace the module:
1.
Turn off power.
2.
Remove plastic cover.
3.
Disconnect connectors.
4.
Remove nuts holding Power Supply.
5.
Remove Power Supply Module.
6.
Replace Power Supply Module.
7.
Reverse steps for installation.
8.
(Refer to Installation Drawings for proper connector
assembly.)
9.
Recalibrate liquid and span values.