Total Organic Carbon Analyzer
Setup and Operation
Teledyne Analytical Instruments
39
Note: On the initial Startup, IR gas calibration should be
performed prior to liquid calibration
a.
Select [CALIBRATE] from the main menu. The
following screen will appear:
b.
Select [TOC].
The following screen appears:
c.
Flow “ZERO” (D. I. Water) and allow NDIR reading
to stabilize. For maximum accuracy, especially at
lower ranges, this process may take 20 minutes.
When reading is stable, select [ZERO].
d.
Flow Liquid Span Standard Solution. Allow NDIR
reading to stabilize, (approximately 15 minutes).
e.
Select [SPAN]. It is recommended that the operator
continue to observe the reading and if not stabilized at
the desired span setting ± 2%, repeat the span
selection until stable.