Theory of Operation
Model 6750
Teledyne Analytical Instruments
16
The construction of the patented UV reactor in the Model 6750
affords superior oxidation by precisely controlling the sample
temperature, adding a precise amount of ultra-pure persulfate reagent
(P/N ST40000) and precisely controlling the carrier gas flow rate.
Pump (P-1) continuously meters a precise amount of carbonate-free
sample from the sparger to the UV reactor. Pump (P-3) continuously
meters a precise amount of persulfate reagent to the reactor. Mass Flow
Controller (MFC-1) precisely controls the carrier flow rate. Refer to
Figures 2-6 and 2-7.
Figure 5-10 shows UV oxidation methodology.
Figure 2-9: Oxidation Reactions