WavePro HD Getting Started Guide
39
Cleaning
Clean the outside of the WavePro HD using a soft cloth moistened
with water or isopropyl alcohol solution. Do not use harsh or abrasive
cleansers. Dry thoroughly before using. Do not submerge the instrument
or allow moisture to penetrate it.
Activating Software Options
To purchase an option (p.42), contact your Teledyne LeCroy sales
representative. You will receive a license key via email that activates the
optional features on the oscilloscope. To install the key on the oscilloscope:
1. From the menu bar, choose
Utilities > Utilities Setup > Options
.
2. Touch
Add Key
.
3. Enter the new
license key
and click
OK
.
4. Reboot the instrument.
Calibration
The WavePro HD is calibrated at the factory prior to being shipped. The
calibration is run at 23 °C (± 2 °C) and is valid for temperatures ± 5 °C of
the original calibration temperature. Within this temperature range, the
WavePro HD will meet all specifications once warmed up.
Warm up the WavePro HD for at least 20 minutes prior to each use.
You will see a message on the oscilloscope that it is still warming up.
Specifications are not guaranteed during the warm up period.
Dynamic calibrations may occur during operation to maintain specified
performance. These are managed by the software in such a way as to
cause least disruption to your work.
We recommend manual calibration if using the oscilloscope in an
environment more than ± 5 °C from the calibration temperature, or if it has
been over one month since the previous calibration. From the menu bar,
choose
Utilities > Preference Setup > Calibration
. The Calibration dialog
will tell you if calibration is needed. There are two routines for selection:
Calibrate All
calibrates all possible combinations of vertical and horizontal
settings at the current environmental conditions. This calibration is valid
for the current temperature ± 5 °C and may take over 60 minutes to
complete.
Calibrate Current Setting
calibrates the current vertical and horizontal
setting at the current environmental conditions. This calibration is valid
at this setting only for the current temperature ± 5 °C and takes about 10
seconds to complete.
CAUTION
. Remove all inputs before calibrating the oscilloscope.
Schedule factory calibration once per year. Contact your local Teledyne
LeCroy office for service.
MAINTENANCE