5.2. FRONT PANEL MENUS
The monitor has an interactive menu structure that can be operated from the front
panel. From within this menu structure, the operator can view real-time parameters
such as temperature and pressure, view configuration information, and edit setup
parameters such as alarm limits. Section 5.1 describes the control keys and
navigation through the menu system. Table 5-1 below shows the menu structure.
Table 5-1. Menu Structure
Sensor
(scroll display to view detector measure values (mV), detector reference values
(mV), cell temperature and cell pressure readings)
Diag
Analog Step
As a way to test connected equipment, this mode allows the user
to step through five analog output V or I values that represent the
ozone concentration range (Section 5.2.3.1)
Cal
Zero Cal (run zero air 10 min. until stable)
Span Cal (use known conc of O3; set target (O3 units set in
Vars); allow span gas to enter for 1 hour prior to execute)
Flow Cal (set flow target based on external flow meter; execute)
Alarm
(screen becomes Default Screen when alarm is triggered; solid-fill indicates which
alarm(s) triggered. Reset allows user to clear an alarm if the ALARM MODE
setting is set to LATCHING ON
Setup
Streams
Allows user to set up Alarm mode and Alarm configurations
Stream duration and Range
Vars
(editable Vars)
Comm
Allows user to set MODBUS ID and View Ethernet Info and IP
Mode
Info
Version information of hardware and firmware
Factory
Default*
Restores the unit to the factory calibration settings and erases
any previously input user settings.
*Do not use this menu without explicit direction by Technical
Support.
5.2.1. EDITING SETTINGS
Refer to Section 5.1.2 for control key descriptions. After navigating to and selecting
a menu item, edit the settings by using the up arrow or down arrow. Either accept
the new setting by pressing the Enter button or exit without changing the setting by
pressing the Back button.
Note that limit checking is enforced while editing values and changes that would
result in invalid values are ignored.
Содержание 480M
Страница 2: ...THIS PAGE IS INTENTIONALLY LEFT BLANK...
Страница 6: ...iv Teledyne API Model 480M O3 Monitor User Manual 09420A DCN8343 This page intentionally left blank...
Страница 12: ...10 Teledyne API Model 480M O3 Monitor User Manual 09420A DCN8343 This page intentionally left blank...
Страница 17: ...09420A DCN8343 Teledyne API Model 480M O3 Monitor User Manual 15...
Страница 21: ...09420A DCN8343 Teledyne API Model 480M O3 Monitor User Manual 19 Figure 3 4 Interior Location of I O Connectors...
Страница 30: ...28 Teledyne API Model 480M O3 Monitor User Manual 09420A DCN8343 This page intentionally left blank...
Страница 41: ...09420A DCN8343 Teledyne API Model 480M O3 Monitor User Manual 39 Figure 6 1 RS 232 RS 485 Signal Configurations...
Страница 54: ...52 Teledyne API Model 480M O3 Monitor User Manual 09420A DCN8343 This page intentionally left blank...