Table 3-3. Status Output States - Failsafe Operation ON, No Error State Detected
Status Output
Failsafe Operating State
Transistor Status
Sensor OK *
Active
Invalid Reading
Active
Check Lamp
Active
Pneumatic Error
Active
*
Failsafe setting does not change Sensor OK behavior.
Table 3-5. Relay Functions - Fail Safe Operation ON, No Alarms Triggered
Relay #
Function
Fail Safe Operating State
1
System OK * Energized
2
Hi Alarm
Energized
3
Hi-Hi Alarm
Energized
*
Failsafe setting does not change System OK behavior.
3.4.5. DIGITAL COMMUNICATION INTERFACES
Note
Ensure that there are no communication devices
connected until after start-up is complete.
3.4.4.1 E
THERNET
For network or Internet communication with the monitor, connect an Ethernet cable
from the Ethernet interface connector (see Figure 3-4) to an Ethernet port. Although
the default setting is DHCP-enabled, it should be manually configured with a Static
IP address (Section 6.1).
3.4.4.2 S
ERIAL
:
RS-232
OR
RS-486
For Serial communication, connect a 9-pin serial cable from the serial connector J14
(see Figure 3-4) to a personal computer serial port. If communication cannot be
established, it’s possible that the cable is pinned differently, and the signal pinout
for the RS-232 connector can be swapped. See Section 6.2 for additional details.
Содержание 480M
Страница 2: ...THIS PAGE IS INTENTIONALLY LEFT BLANK...
Страница 6: ...iv Teledyne API Model 480M O3 Monitor User Manual 09420A DCN8343 This page intentionally left blank...
Страница 12: ...10 Teledyne API Model 480M O3 Monitor User Manual 09420A DCN8343 This page intentionally left blank...
Страница 17: ...09420A DCN8343 Teledyne API Model 480M O3 Monitor User Manual 15...
Страница 21: ...09420A DCN8343 Teledyne API Model 480M O3 Monitor User Manual 19 Figure 3 4 Interior Location of I O Connectors...
Страница 30: ...28 Teledyne API Model 480M O3 Monitor User Manual 09420A DCN8343 This page intentionally left blank...
Страница 41: ...09420A DCN8343 Teledyne API Model 480M O3 Monitor User Manual 39 Figure 6 1 RS 232 RS 485 Signal Configurations...
Страница 54: ...52 Teledyne API Model 480M O3 Monitor User Manual 09420A DCN8343 This page intentionally left blank...