7.3. PARTICULATE FILTER REPLACEMENT
1.
Disconnect power from the monitor.
2.
Locate the particulate housing filter on the bottom panel and loosen the four screws
on the sample filter body (see Figure 7-2).
3.
Remove the fours screws, the four washers and the sample filter body from the
sample filter base.
4.
Remove the o-ring and inspect for any cracks or deformities. If there are cracks or
deformities, replace the o-ring with a new one; otherwise, it can be reused.
5.
Remove the two sample filter retainers and glass fiber filter element from the sample
filter body.
6.
Discard the glass fiber filter element.
7.
Place a new glass fiber filter element in between the two sample filter retainers and
place in the sample filter body.
8.
Place the sample filter body on the sample body base and secure with the four screws
and four washers.
9.
Close the front panel.
Figure 7-2. Particulate Filter (PN 02832) Replacement
Содержание 480M
Страница 2: ...THIS PAGE IS INTENTIONALLY LEFT BLANK...
Страница 6: ...iv Teledyne API Model 480M O3 Monitor User Manual 09420A DCN8343 This page intentionally left blank...
Страница 12: ...10 Teledyne API Model 480M O3 Monitor User Manual 09420A DCN8343 This page intentionally left blank...
Страница 17: ...09420A DCN8343 Teledyne API Model 480M O3 Monitor User Manual 15...
Страница 21: ...09420A DCN8343 Teledyne API Model 480M O3 Monitor User Manual 19 Figure 3 4 Interior Location of I O Connectors...
Страница 30: ...28 Teledyne API Model 480M O3 Monitor User Manual 09420A DCN8343 This page intentionally left blank...
Страница 41: ...09420A DCN8343 Teledyne API Model 480M O3 Monitor User Manual 39 Figure 6 1 RS 232 RS 485 Signal Configurations...
Страница 54: ...52 Teledyne API Model 480M O3 Monitor User Manual 09420A DCN8343 This page intentionally left blank...