6. DIGITAL COMMUNICATIONS
The 480M comes equipped with digital communications capability that can be
connected to a computer or digital data acquisition system (DAS). See Section 3.4.4
for configuration information. There is a serial communications connector that can
be configured for RS-232 or RS485 and an Ethernet connector that uses Ethernet 10
Mbit standard. All three use the standardized MODBUS protocol (see Section 6.3
for register maps).
The connectors can be used for data acquisition, alarm triggering, and instrument
configuration. All the functions that are available at the front panel of the instrument
can also be performed over the digital communications standards.
6.1. ETHERNET CONFIGURATION
Once an Ethernet cable is connected (Section 3.4.4.1), the instrument should be
functioning with a dynamic IP address by default. To configure with a static
connection, use the front panel control keys (Section 5.1.2) to navigate to the
Setup>Comm>Ethernet>IP mode menu. If you don’t know what the IP Address,
Mask, or Gateway values should be, consult with your IT Administrator first.
1.
Press the Enter button to display the menu.
2.
Press the applicable Select button to navigate to the next menu page.
3.
Press the Select button under Setup.
4.
Press the Select button under Comm.
5.
Press Select button under IP mode.
6.
Press the Select button to highlight “Static” and press Enter to accept.
7.
Press the Back button to move to the previous level.
8.
Press the Select button under IP Addr.
9.
Press the Select buttons under the right/left arrows as needed to highlight the
first place of the IP address string.
10.
Press the Select buttons under the up/down (increment/decrement) arrows as
needed to change the value in the highlighted place of the IP address string.
11.
Repeat the navigation and selection steps until the address is complete.
12.
Press the Enter button to accept and the Back button to move to the previous
level.
13.
Repeat the process for Mask and Gateway if/as directed by IT Admin.
Содержание 480M
Страница 2: ...THIS PAGE IS INTENTIONALLY LEFT BLANK...
Страница 6: ...iv Teledyne API Model 480M O3 Monitor User Manual 09420A DCN8343 This page intentionally left blank...
Страница 12: ...10 Teledyne API Model 480M O3 Monitor User Manual 09420A DCN8343 This page intentionally left blank...
Страница 17: ...09420A DCN8343 Teledyne API Model 480M O3 Monitor User Manual 15...
Страница 21: ...09420A DCN8343 Teledyne API Model 480M O3 Monitor User Manual 19 Figure 3 4 Interior Location of I O Connectors...
Страница 30: ...28 Teledyne API Model 480M O3 Monitor User Manual 09420A DCN8343 This page intentionally left blank...
Страница 41: ...09420A DCN8343 Teledyne API Model 480M O3 Monitor User Manual 39 Figure 6 1 RS 232 RS 485 Signal Configurations...
Страница 54: ...52 Teledyne API Model 480M O3 Monitor User Manual 09420A DCN8343 This page intentionally left blank...