3.3.2. WIRING INSTRUCTIONS
1.
Install a ½” conduit fitting for routing the electrical wiring into the monitor
through the hole provided in the bottom face of the NEMA enclosure; an
appropriate sealed conduit connector should be used.
2.
Hard-wire the electrical connection to the terminal block and grounding lug
on the chassis. AC power connection to the monitor should be made with 12-
14 AWG stranded copper wire, connected to the monitor as follows:
•
Earth Ground (green):
Connect the earth ground wire to the copper grounding lug
•
Line (Black):
Connect the Line wire to the terminal block directly across
from the black wire that leads to the power supply
•
Neutral (White):
Connect the Neutral wire to the terminal block directly
across from the white wire that leads to the power supply.
Figure 3-3. AC Power Connection to Monitor
Содержание 480M
Страница 2: ...THIS PAGE IS INTENTIONALLY LEFT BLANK...
Страница 6: ...iv Teledyne API Model 480M O3 Monitor User Manual 09420A DCN8343 This page intentionally left blank...
Страница 12: ...10 Teledyne API Model 480M O3 Monitor User Manual 09420A DCN8343 This page intentionally left blank...
Страница 17: ...09420A DCN8343 Teledyne API Model 480M O3 Monitor User Manual 15...
Страница 21: ...09420A DCN8343 Teledyne API Model 480M O3 Monitor User Manual 19 Figure 3 4 Interior Location of I O Connectors...
Страница 30: ...28 Teledyne API Model 480M O3 Monitor User Manual 09420A DCN8343 This page intentionally left blank...
Страница 41: ...09420A DCN8343 Teledyne API Model 480M O3 Monitor User Manual 39 Figure 6 1 RS 232 RS 485 Signal Configurations...
Страница 54: ...52 Teledyne API Model 480M O3 Monitor User Manual 09420A DCN8343 This page intentionally left blank...