Merging Modules
Figure 53: Feeding the merge cable through the cover
10.
Turn the merge cable bracket over so that the guide pins point up.
11.
Place the bracket over the merge cable connector.
12.
Install the two screws that hold the merge cable bracket in place.
CAUTION.
To prevent damage to the module during the installation process,
reinstall the cover exactly as described in steps 14 through 19.
If the cover is not properly seated, the module can be damaged when you install it
in the mainframe and it may not meet EMC requirements.
13.
Replace the logic analyzer module cover.
14.
Push forward on the cover so the tab on the front edge of the cover inserts
into the rear of the front subpanel. Make sure that the cover is fully seated
(no gaps) against the front and rear chassis
fl
anges.