Adjustment Procedures
5- 8
TLA Series Digitizing Oscilloscope Modules Service Manual
This procedure adjusts the frequency response of the TLA7E1 and the TLA7E2
DSO modules.
SW test names
BANDWIDTH_CAL
Equipment
required
Sine wave generator
1
(item 14)
Precision BNC cable (item 12)
Type N-to-BNC adapter (item 16)
Prerequisites
Warm-up time: 30 minutes
The logic analyzer application is not running
Test equipment connected as shown in Figure 5--2
2
Power-up diagnostics pass
Base Calibration completed
1
The output frequency of the sine wave generator must maintain a constant output
amplitude within 0.35 dB of the reference frequency (10 MHz) through the bandwidth
calibration frequency range (1005 MHz).
2
If you require a power meter to maintain a constant output amplitude use the
equipment setup as shown in Figure 5- 3.
Sine wave generator
Type N-to-BNC
adapter
DSO module
CH 1
Precision
BNC cable
Figure 5- 2: Initial setup for the frequency response adjustment
DSO Procedure 2:
Frequency Response
Adjustment
Содержание TLA Series
Страница 4: ......
Страница 12: ...Service Safety Summary viii TLA Series Digitizing Oscilloscope Modules Service Manual...
Страница 16: ...Preface xii TLA Series Digitizing Oscilloscope Modules Service Manual...
Страница 20: ...Introduction xvi TLA Series Digitizing Oscilloscope Modules Service Manual...
Страница 40: ...Theory of Operation 3 4 TLA Series Digitizing Oscilloscope Modules Service Manual...
Страница 114: ...Repackaging Instructions 6 36 TLA Series Digitizing Oscilloscope Modules Service Manual...
Страница 116: ...Options 7 2 TLA Series Digitizing Oscilloscope Modules Service Manual...
Страница 118: ...Replaceable Electrical Parts 8 2 TLA Series Digitizing Oscilloscope Modules Service Manual...
Страница 129: ......
Страница 130: ......