Operating Information
2- 2
TLA Series Digitizing Oscilloscope Modules Service Manual
NOTE
. Do not set the digitizing oscilloscope logical address to 00. Logical
address 00 is reserved for the controller.
Most-significant
digit
Least-significant
digit
Figure 2- 1: Logical address switches
Dynamic Auto Configuration.
With Dynamic Auto Configuration selected
(hexadecimal FF or decimal 255), the system automatically sets the address to an
unused value. For example, if there are devices set to addresses 01 and 02
already in your system, the resource manager will automatically assign the DSO
module an address other than 01 or 02.
Static Logical Address.
Static logical address selections set the address to a fixed
value. If you use static logical addressing, each device within your system must
have a unique address to avoid communication problems.
Содержание TLA Series
Страница 4: ......
Страница 12: ...Service Safety Summary viii TLA Series Digitizing Oscilloscope Modules Service Manual...
Страница 16: ...Preface xii TLA Series Digitizing Oscilloscope Modules Service Manual...
Страница 20: ...Introduction xvi TLA Series Digitizing Oscilloscope Modules Service Manual...
Страница 40: ...Theory of Operation 3 4 TLA Series Digitizing Oscilloscope Modules Service Manual...
Страница 114: ...Repackaging Instructions 6 36 TLA Series Digitizing Oscilloscope Modules Service Manual...
Страница 116: ...Options 7 2 TLA Series Digitizing Oscilloscope Modules Service Manual...
Страница 118: ...Replaceable Electrical Parts 8 2 TLA Series Digitizing Oscilloscope Modules Service Manual...
Страница 129: ......
Страница 130: ......