![Tektronix TLA Series Скачать руководство пользователя страница 55](http://html1.mh-extra.com/html/tektronix/tla-series/tla-series_service-manual_1078206055.webp)
Performance Verification
TLA Series Digitizing Oscilloscope Modules Service Manual
4- 15
This procedure checks the offset accuracy of the DSO module.
SW test name
OFFSET_ACCURACY
Equipment
required
No change from the DC Gain Accuracy test
Prerequisites
Warm-up time: 30 minutes
The logic analyzer application is not running
The performance verification and adjustment software is loaded
Test equipment connected as shown in Figure 4--3
Power-up diagnostics pass
INTERNAL_CAL test passes
1.
Verify that all of the prerequisites listed previously are met for the procedure.
WARNING.
These procedures use voltage levels that step through
±
99.90 volts.
Use caution when using voltages approaching this magnitude.
2.
Follow the on-screen instructions to run the test for each channel on the
DSO module.
3.
After the test begins, make changes to test equipment settings and connec-
tions when requested by the on-screen instructions.
4.
Verify that all of the tests pass. If a test fails, check the
Troubleshooting
section on page 4--12.
DSO Procedure 3:
Offset Accuracy
Содержание TLA Series
Страница 4: ......
Страница 12: ...Service Safety Summary viii TLA Series Digitizing Oscilloscope Modules Service Manual...
Страница 16: ...Preface xii TLA Series Digitizing Oscilloscope Modules Service Manual...
Страница 20: ...Introduction xvi TLA Series Digitizing Oscilloscope Modules Service Manual...
Страница 40: ...Theory of Operation 3 4 TLA Series Digitizing Oscilloscope Modules Service Manual...
Страница 114: ...Repackaging Instructions 6 36 TLA Series Digitizing Oscilloscope Modules Service Manual...
Страница 116: ...Options 7 2 TLA Series Digitizing Oscilloscope Modules Service Manual...
Страница 118: ...Replaceable Electrical Parts 8 2 TLA Series Digitizing Oscilloscope Modules Service Manual...
Страница 129: ......
Страница 130: ......