Performance Verification
TLA Series Digitizing Oscilloscope Modules Service Manual
4- 3
Test Equipment
These procedures use external, traceable signal sources to directly test character-
istics that are designated as checked (
n
) in the
Specifications
chapter of this
manual. Table 4--1 shows the required equipment list. Always warm up the
equipment for 30 minutes before beginning the procedures.
Table 4- 1: Test Equipment
Item number and description
Minimum requirements
Example
1. Mainframe
TLA700 Series Mainframe with a DSO
module installed
TLA721 Benchtop mainframe or
TLA715 Portable Mainframe
2. DSO probe calibration fixture
One required
Tektronix part number 671-3930-XX
3. DSO probes
Two required, with < one-inch ground leads
Tektronix P6243 or P6245 probe, with std.
accessories
4. Adapter, male BNC-to-male BNC
One required
Tektronix part number 103-0029-XX
5. Digital multimeter with test leads
DCV accuracy: 0.1% from --10 V to +100 V
Tektronix DMM 900 Series
6. Connector, dual-banana
Female BNC-to-dual banana
Tektronix part number 103-0090-XX
7. Voltage reference
Accuracy:
≤
0.01%
Data Precision 8200
8. Capacitor
1
0.1
F, 200 V
Tektronix part number 283-0189-XX
9. Cable, dual-input
Female BNC-to-dual male BNC
Tektronix part number 067-0525-XX
10. Adapter, N-to-BNC
Male type N-to-female BNC
Tektronix part number 103-0045-XX
11. Adapter, SMA-to-BNC
Female type SMA-to-female BNC
Tektronix part number 015-1018-XX
12. Cable, precision 50
Ω
coaxial
50
Ω
, 36 in, male-to-male BNC connectors
Tektronix part number 012-0482-XX
13. Generator, function
Frequency range: 1 Hz to 20 MHz
Frequency accuracy: 0.1%
Amplitude range: 10 mV to 20 V
p-p
Hewlett Packard 3325B
14. Generator, sine wave
Frequency range: 100 kHz to 1005 MHz
Frequency accuracy: <6 ppm
Amplitude: 2 mV to 1.5 V
RMS
Accuracy:
<0.35 dB
Gigatronics 6061
15. Power meter with sensor
Bandwidth: >1.2 GHz Accuracy:
0.2 dB Sensitivity: 500 pW to 20 mW
Rohde & Schwarz NRVS with model
NRV-Z4 Sensor
16. Adapter, N-to-BNC (used only with
power meter)
Female type N-to-male BNC
Tektronix part number 103-0058-XX
17. 50
Ω
power divider (used only with
power meter)
Provide load isolation between equipment
Maximum VSWR: 1.50
Tektronix part number 015-0565-XX
18. SMA-to-BNC adapters (used only with
power meter)
For use with the 50
Ω
power divider
(three required)
Tektronix part number 015-1018-XX
19. Printer with cable (optional, to print
Calibration Data Reports)
PC-compatible, continuous feed, prints
ASCII text, connects to EPP connector
Any general purpose printer
1
The capacitor is installed across the Data Precision 8200 output terminals to reduce noise. If your voltage reference
produces <4 mVp-p of noise, external noise reduction is not necessary.
Содержание TLA Series
Страница 4: ......
Страница 12: ...Service Safety Summary viii TLA Series Digitizing Oscilloscope Modules Service Manual...
Страница 16: ...Preface xii TLA Series Digitizing Oscilloscope Modules Service Manual...
Страница 20: ...Introduction xvi TLA Series Digitizing Oscilloscope Modules Service Manual...
Страница 40: ...Theory of Operation 3 4 TLA Series Digitizing Oscilloscope Modules Service Manual...
Страница 114: ...Repackaging Instructions 6 36 TLA Series Digitizing Oscilloscope Modules Service Manual...
Страница 116: ...Options 7 2 TLA Series Digitizing Oscilloscope Modules Service Manual...
Страница 118: ...Replaceable Electrical Parts 8 2 TLA Series Digitizing Oscilloscope Modules Service Manual...
Страница 129: ......
Страница 130: ......