Performance Verification
4- 2
TLA Series Digitizing Oscilloscope Modules Service Manual
Determine equipment
to certify.
Go to appropriate
procedure(s).
Gather required test
equipment.
Use
Tektronix-supplied
probes.
Configure customer’s
mainframe and modules as
described in procedure.
Set up test equipment.
Select procedure.
Allow all mainframes, modules and
equipment to warm up (30 minutes).
Are
all certifications
complete?
Done
Is it OK
with customer to
remove/reconnect probes
from System
Under Test?
No
No
Yes
Do tests
pass?
Repair or adjust
equipment as
required.
No
Yes
Yes
Perform procedure
using software.
Load PV/Adjust
software.
Calibration
Data Report.
Create necessary
Calibration Certificate.
Figure 4- 1: Calibration/certification procedure flow chart
Содержание TLA Series
Страница 4: ......
Страница 12: ...Service Safety Summary viii TLA Series Digitizing Oscilloscope Modules Service Manual...
Страница 16: ...Preface xii TLA Series Digitizing Oscilloscope Modules Service Manual...
Страница 20: ...Introduction xvi TLA Series Digitizing Oscilloscope Modules Service Manual...
Страница 40: ...Theory of Operation 3 4 TLA Series Digitizing Oscilloscope Modules Service Manual...
Страница 114: ...Repackaging Instructions 6 36 TLA Series Digitizing Oscilloscope Modules Service Manual...
Страница 116: ...Options 7 2 TLA Series Digitizing Oscilloscope Modules Service Manual...
Страница 118: ...Replaceable Electrical Parts 8 2 TLA Series Digitizing Oscilloscope Modules Service Manual...
Страница 129: ......
Страница 130: ......