![Tektronix TLA Series Скачать руководство пользователя страница 68](http://html1.mh-extra.com/html/tektronix/tla-series/tla-series_service-manual_1078206068.webp)
Adjustment Procedures
5- 4
TLA Series Digitizing Oscilloscope Modules Service Manual
Adjustment Instructions
This section describes how to perform adjustments using the performance
verification and adjustment software.
The performance verification and adjustment software contains instructions for
performing the adjustments. The basic steps for completing the procedures
follow:
1.
Start the program and enter user and product identification information.
2.
Select a full adjustment sequence.
3.
Connect the test equipment.
4.
Run each adjustment step as instructed.
5.
After completing all the adjustment steps, view the test results to confirm
that the adjustment was successful.
When a test passes, the software automatically loads new calibration data into
the DSO module memory.
NOTE
. Use the dual input cable or a set of matched cables (for electrical length)
when running the CHAN_SKEW_CAL test.
If you use unmatched separate coaxial cables, the adjustment may fail or may be
miscalibrated due to an electrical mismatch between the cables.
Using the PV/Adjust
Software
Содержание TLA Series
Страница 4: ......
Страница 12: ...Service Safety Summary viii TLA Series Digitizing Oscilloscope Modules Service Manual...
Страница 16: ...Preface xii TLA Series Digitizing Oscilloscope Modules Service Manual...
Страница 20: ...Introduction xvi TLA Series Digitizing Oscilloscope Modules Service Manual...
Страница 40: ...Theory of Operation 3 4 TLA Series Digitizing Oscilloscope Modules Service Manual...
Страница 114: ...Repackaging Instructions 6 36 TLA Series Digitizing Oscilloscope Modules Service Manual...
Страница 116: ...Options 7 2 TLA Series Digitizing Oscilloscope Modules Service Manual...
Страница 118: ...Replaceable Electrical Parts 8 2 TLA Series Digitizing Oscilloscope Modules Service Manual...
Страница 129: ......
Страница 130: ......