Access Shortcut Menus by Right-Clicking
—fast access to most common commands is available by right-clicking
Right-click on the graticule, on an object, or a readout for quick access to shortcut menus. These
shortcut menus are dynamic and will change with the area or object where you right clicked.
Right-click on any of the following for a dynamic shortcut menu:
Waveform handle
Cursors
Labels
Histogram display
Masks
Templates
Trigger level marker
Readouts (the menu varies with the type of readout)
Graticule
Regions (Menu bar, Toolbar, Multipurpose Knobs)
Initializing the Instrument
—returning to a known starting point
After you complete a task, you may want to initialize the instrument settings before using the
instrument with your next task. This may be especially important if you share the instrument with
others; previous settings may make it difficult for you to do what you need to do.
To quickly restore the instrument to its default settings, push the front-panel DEFAULT SETUP
button.
If data security is important to you, use the Tek Secure Erase feature to overwrite all internal
waveforms and set up memory with null values.
Choosing and Connecting Probes
—important things to remember about probes
Choose from a wide variety of passive, active, and specialty probes to use with your instrument.
The instrument automatically configures the input impedance when you attach a probe that is
compatible with the TekProbe interface. Remember the following when connecting a probe:
Adjust the low frequency compensation on most passive probes after you connect them to a
channel input.
Compensate for gain (and offset) error in many probes. Connect the probe tip to the probe
compensation connectors and execute the built-in calibration procedure to optimize the
system accuracy.
Compensate for propagation delay differences between voltage probes using the probe
deskew feature. Attach up to two probe tips at a time to the probe compensation terminals and
execute the deskew procedure to time align the display for each channel. The deskew
adjustment range is
±
75 ns.