193
Limit Test Failure Notification Control Window
Access this control window from the Mask menu, select Limit Test Setup…, and then click Config.
Overview
Use the Failure Notification control window to set the mode of notification, send commands, save
waveforms, print, send email and log data when a limit test fails.
To Use
Click the On/Off buttons to turn that mode of notification on or off.
Click Path to configure the path for saving waveforms and the Log Date.
Click Config to open the E-mail on Event Setup control window.
Behavior
When a limit test fails and Beep is turned on, the instrument beeps, when SRQ is turned on, it
sends a GPIB SRQ command, and when Stop ACQ is turned on, the instrument stops acquiring.
When Print is turned on and a limit test fails, the instrument sends a screen print to the default
printer. The current page setup is used for the screen shot. There is no specific page setup for limit
test failures.
When Save Wfm is turned on and a limit test fails, the instrument saves waveform data to a
default directory unless you specified a different directory using the Path button. A file name is
generated by combining the date and time.
When Log Date is turned on and a limit test fails, the date and time for each failed limit test is
recorded to a file of your choice.
When E-mail is turned on and a limit test fails, the instrument sends an e-mail message to the
specified recipient(s). You must enter an e-mail address and the SMTP server name in the E-mail
on Event Setup control window before you can turn on E-mail. Clicking Config opens the E-mail
on Event Setup control window where you can enter the e-mail address(es) and an SMTP server
name. Check with your system administrator for the correct server name.
Содержание TDS5000B Series
Страница 36: ...36 AFTDS Differential Signal Adapter...
Страница 280: ...280 Blackman Harris window...