339
horizontal and vertical divisions), return to default settings,
or return to the Mask Setup control window. Use the keypad
to change the vertical or horizontal autofit parameters.
11.
Click the Masks button to return to the Mask control
window.
Set mask test tolerance
margins
12.
Click the Masks Setup window Tolerance tab to set the
percentage of margin used in the mask test. Use the control
knob, keypad, pop-up keypad, or up and down arrow
buttons to enter the mask margin tolerance percentage. The
range of values is –50% to 50%.
Margin tolerance settings greater than 0% expand the size of the
segments, making the mask test harder to pass; margin tolerance
settings less than 0% (negative percent) reduces the size of the
segments, making the mask test easier to pass.
Set mask test pass and
fail parameters
13.
Select the Pass/Fail Setup tab of the Masks control
window.
14.
Use the control knob, keypad, or pop-up keypad to enter
the number of waveforms to test (number of samples in
some modes), the failure threshold (the number of
waveforms that must fail to fail the test), and the delay
time (the time from when mask test starts to when the
instrument begins sampling).
15.
Use the Failure field buttons to set what the instrument
does when a mask test fails; have the instrument beep
(BEEP), send an SRQ out on the GPIB bus (SRQ), stop
signal acquisition immediately (Stop Acq), send email (E-
mail), and/or print the instrument screen image to a printer
(Print).
You many need to select a sound source to enable the BEEP. See
Selecting the Sound Source on page 346 for more information.
16.
Clicking the More button displays more failure functions.
Save Wfm saves the the waveform data of the first
waveform that causes the test to fail to a .wfm file. Log
Date saves time, date, and basic test information of the
first waveform that causes the test to fail to an ASCII text
(.txt) file.
Both files are saved to the location specified by the Path button.
The file name format is YYMMDD-HHMMSS, where YY is
year, MM is month, DD is day, HH is hour, MM is minutes, and
SS is seconds.
17.
Use the Completion field buttons to set what the
instrument does at the completion of a mask test.
18.
Use the Polarity buttons to set mask and waveform
polarity. Positive tests the positive waveform pulses.
Negative inverts the mask and tests the negative waveform