Command Descriptions
TDS Family Oscilloscope Programmer Manual
2–173
HORizontal:MAIn:SCAle <NR3>
HORizontal:MAIn:SCAle?
:
SCAle
HORizontal
:
MAIn
?
<Space>
<NR3>
<NR3>
is the time per division. For the TDS 400A series, the range is 20 s to
1 ns. For the TDS 510A, 500B, 600B, and 700A, the range is 10 s to 500 ps (or
200 ps on the TDS 784A), in a 1–2–5 sequence.
In extended-acquisition-length mode, no matter what value you set, the
oscilloscope will only use the closest real-time, non-interpolated value.
HORIZONTAL:MAIN:SCALE 2E-6
sets the main scale to 2
m
s per division.
HORizontal:MAIn:SECdiv
Sets the time per division for the main time base. This command is identical to
the HORizontal:MAIn:SCAle command. It is provided to maintain program
compatibility with some older models of Tektronix digitizing oscilloscopes.
Horizontal
HORizontal:DELay:SCAle, HORizontal:DELay:SECdiv, HORizon-
tal:MAIn:SCAle
HORizontal:MAIn:SECdiv <NR3>
HORizontal:MAIn:SECdiv?
HORizontal
:
SECdiv
:
MAIn
?
<Space>
<NR3>
HORizontal:MODe
Selects whether the horizontal display uses the main or delayed time base or
both. This command is equivalent to setting Time Base in the Horizontal menu.
Syntax
Arguments
Examples
Group
Related Commands
Syntax