Command Descriptions
TDS Family Oscilloscope Programmer Manual
2–51
ACQuire:NUMACq?
ACQuire
:
NUMACq
?
<NR1>
ACQUIRE:NUMACQ?
might return
350
, indicating that 350 acquisitions took place since an AC-
QUIRE:STATE RUN command was executed.
ACQuire:NUMAVg
Sets the number of waveform acquisitions that make up an averaged waveform.
This is equivalent to setting the Average count in the Acquisition Mode side
menu.
Acquisition
ACQuire:MODe
ACQuire:NUMAVg <NR1>
ACQuire:NUMAVg?
NUMAVg
?
<Space>
<NR1>
:
ACQuire
<NR1>
is the number of waveform acquisitions, from 2 to 10,000.
ACQUIRE:NUMAVG 10
specifies that an averaged waveform will show the result of combining 10
separately acquired waveforms.
ACQUIRE:NUMAVG?
might return
75
, indicating that there are 75 acquisitions specified for averaging.
Syntax
Related Commands
Examples
Group
Related Commands
Syntax
Arguments
Examples