Command Descriptions
TDS Family Oscilloscope Programmer Manual
2–287
1 msec/division then the default holdoff will be 1 msec/division
×
25 divs =
25 msec.
sets the holdoff to the by time setting. This enables the user to set the holdoff
time.
TRIGger:MAIn:HOLDOff:TIMe
TDS 500C, 600B & 700C Only
Sets or queries the main trigger holdoff time. This is equivalent to setting
Holdoff Time in the Mode & Holdoff side menu.
Trigger
holdoff time in seconds. The range is 250 ns to 12.0 seconds.
sets the holdoff time to be 10 seconds.
TRIGger:MAIn:HOLDOff:VALue
TDS 400A & 510A Only
Sets or queries the main trigger holdoff value. This is equivalent to setting
Holdoff in the Mode & Holdoff side menu.
Trigger
Examples
Group
Syntax
Arguments
Examples
Group