Appendix D: Factory Initialization Settings
TDS Family Oscilloscope Programmer Manual
D–5
Table D–1: Factory Initialization Settings (Cont.)
Control
Changed by Factory Init to
Measure gating
Off
Measure high ref
90% and 0 V (units)
Measure high-low setup
Histogram
Measure low ref
10% and 0 V (units)
Measure mid ref
50% and 0 V (units)
Measure mid2 ref
50% and 0 V (units)
Pulse glitch filter state
TDS 510A, 500C, 600B & 700C: On
(Accept glitch)
Pulse glitch trigger polarity
TDS 510A, 500C, 600B & 700C: Positive
Pulse glitch width
TDS 510A, 500C, 600B & 700C: 2.0 ns
Pulse runt high threshold
TDS 510A, 500C, 600B & 700C: 1.2 V
Pulse runt low threshold
TDS 510A, 500C, 600B & 700C: 0.8 V
Pulse runt trigger polarity
TDS 510A, 500C, 600B & 700C: Positive
Pulse runt triggers when ...
TDS 510A, 500C, 600B & 700C: Occurs
Pulse runt width
TDS 510A, 500C, 600B & 700C: 2.0 ns
Pulse slew rate delta time
2.0 ns
Pulse slew rate polarity
Positive
Pulse slew rate thresholds
Upper
Lower
Trig if faster than
1.80 V
800 mV
Pulse slew rate triggers when ...
Trig if faster than
Pulse timeout polarity
TDS 500C & 600B: Either
Pulse timeout time
TDS 500C & 600B: 2.0 ns
Pulse trigger class
TDS 510A, 500C, 600B & 700C: Glitch
Pulse trigger level
TDS 510A, 500C, 600B & 700C: 0.0 V
Pulse trigger source
(Glitch, runt, and width, and timeout)
TDS 500C, 600B, & 700C:
Channel 1 (Ch1)
Pulse width lower limit
TDS 510A, 500C, 600B & 700C: 2.0 ns
Pulse width trigger polarity
TDS 510A, 500C, 600B & 700C: Positive
Pulse width trigger when
TDS 510A, 500C, 600B & 700C: Within limits
Pulse width upper limit
TDS 510A, 500C, 600B & 700C: 2.0 ns
Repetitive signal
TDS 500C & 700C: On
RS-232 parameters
No change
Saved setups
No change