Performance Veri
fi
cation
5.
Connect the probe to any channel (1–4)
of the oscilloscope, and then degauss the
probe.
6.
Clamp the current probe around the HF
current loop. Verify that the arrow-shaped
indicator on the probe points away from the
pulse generator.
7.
Record the rise time measurement in the
Test Record.
8.
Calculate the probe bandwidth using the
measured rise time in the following formula:
9.
Record the calculated bandwidth value in
the test record.
End of procedures.