Performance Veri
fi
cation
Set up the Equipment
Use the following procedure to set up and warm up the equipment to test the probe.
1.
Turn on the oscilloscope.
2.
Connect the probe to any channel of the
oscilloscope.
3.
Press the
DEGAUSS
button.
4.
Set the oscilloscope coupling to DC.
5.
Power on the current source and the pulse
generator.
6.
Allow 20 minutes for the equipment to
warm up.
7.
Photocopy the test record and use it to
record the test results. (See page 56.)
50
TCP202A Current Probe Instruction Manual