Accessories and Options
TekVPI Calibration Fixture
This calibration
fi
xture is required to complete
a performance veri
fi
cation and gain accuracy
adjustment procedures on the probe. It
provides power to the probe and routes the
probe output signal out through an SMA
connector on the back of the
fi
xture. The signal
can then be measured with another instrument,
such as a precision DMM, to check and adjust
the gain accuracy of the probe.
Order Tektronix part number 067-1701-xx.
24
TCP202A Current Probe Instruction Manual