Performance Veri
fi
cation
12.
Record the exact measurement of the
DMM as M1.
13.
Set the current source output to -0.50 A.
14.
Record the exact measurement of the
DMM as M2.
15.
Compute the % Error using the measured
amplitude values and the formula shown.
For example, you might measure values of
0.250 V for M1 and –0.255 V for M2. With
an expected output voltage (Ve) of 0.250 V,
compute the % Error as shown.
16.
Record the computed %Error values in the
Test Record.
17.
Disable the calibrator output.
Test current = ±2.500 A (±0.50 A source output x 5 coil turns)
Expected output voltage (Ve) = 0.250 V
Example:
TCP202A Current Probe Instruction Manual
53
Содержание TCP202A
Страница 1: ...x TCP202A 50 MHz 15 A AC DC Current Probe ZZZ Instruction Manual P071300300 071 3003 00...
Страница 2: ......
Страница 6: ......
Страница 10: ...Table of Contents iv TCP202A Current Probe Instruction Manual...
Страница 62: ...Specifications Figure 2 Typical input impedance versus frequency 42 TCP202A Current Probe Instruction Manual...
Страница 63: ...Specifications Figure 3 Maximum peak pulse versus pulse width TCP202A Current Probe Instruction Manual 43...
Страница 65: ...Specifications Figure 4 Probe dimensions TCP202A Current Probe Instruction Manual 45...