![Tektronix TCP0020 Скачать руководство пользователя страница 91](http://html1.mh-extra.com/html/tektronix/tcp0020/tcp0020_instruction-manual_1078185091.webp)
Performance Veri
fi
cation
TCP0020 & TCP2020 Test
6.
Do not clamp the current probe around any
conductor, but make sure the jaw is in the
locked position.
7.
Degauss the probe by pressing the
Degauss
or
Degauss/AutoZero
button.
8.
Connect the 5-turn current loop to the
current source, and then clamp the current
probe around the 5-turn current loop as
shown. The arrow indicator on the probe
should point away from the (+) terminal of
the current source.
9.
Set the current source output to +2.00 A.
10.
Set the DMM to measure DC volts, on
≥
2 volt range (or use Autoset).
11.
Enable the output of the current source.