Performance Veri
fi
cation
TCP0020 Setup
1.
Connect a BNC-to-Dual Banana adapter to
the digital multimeter (DMM) input.
2.
Connect the SMA M-to-BNC F adapter
to the SMA output of the TekVPI
Calibration/Veri
fi
cation adapter.
3.
Connect the BNC cable between
the BNC adapter on the TekVPI
Calibration/Veri
fi
cation adapter and the
BNC adapter attached to the DMM.
4.
Connect the TekVPI Calibration/Veri
fi
cation
adapter to any channel (1–4) of the
oscilloscope.
5.
Connect the probe to the TekVPI
Calibration/Veri
fi
cation adapter.
NOTE.
The adapter is only used to supply
power to the probe; measurements are taken
on the DMM.
66
20 A Current Probes Instruction Manual