The speci
fi
cations in Tables 1 through 5 are valid under the following conditions:
The probe has been calibrated at an ambient temperature of 23 °C ±5 °C.
The probe is connected to a host instrument with an input impedance of 1 M
Ω
. The TCP0020 probe speci
fi
cations are also
valid on host instruments with an input impedance of 50
Ω
.
The probe must have a warm-up period of at least 20 minutes and be in an environment that does not exceed the limits
described. (See Table 1).
The probe has been degaussed successfully.
Speci
fi
cations for the TCP0020 & TCP2020 current probes fall into three categories: warranted, typical, and nominal characteristics.
Warranted Characteristics
Warranted characteristics describe guaranteed performance within tolerance limits or certain type-tested requirements. Warranted
characteristics that have checks in the
Performance Veri
fi
cation
section are marked with the
symbol.
Table 1: Warranted electrical characteristics
Characteristic
Description
DC gain accuracy
<3%
Rise time (10% to 90%)
≤
7 ns
Bandwidth
DC to 50 MHz
20 A Current Probes Instruction Manual
53
Содержание TCP0020
Страница 1: ...x TCP0020 TCP2020 50 MHz 20 Amp AC DC Current Probes ZZZ Instruction Manual P071300200 071 3002 00...
Страница 2: ......
Страница 6: ......
Страница 10: ...Table of Contents iv 20 A Current Probes Instruction Manual...
Страница 22: ...Compliance Information xvi 20 A Current Probes Instruction Manual...
Страница 80: ...Specifications Figure 2 Typical insertion impedance versus frequency 56 20 A Current Probes Instruction Manual...
Страница 81: ...Specifications Figure 3 Maximum peak pulse versus pulse width 20 A Current Probes Instruction Manual 57...
Страница 83: ...Specifications Figure 4 Mechanical dimensions 20 A Current Probes Instruction Manual 59...