DC Gain Accuracy
This test checks the DC gain accuracy of the probe. If the measurements are out of the speci
fi
ed limits in the test record, refer to
the
Adjustments
section. (See page 72.)
TCP2020 Setup
1.
Connect a BNC-to-Dual Banana adapter to
the digital multimeter (DMM) input.
2.
Connect the TCP2020 probe output to the
BNC adapter attached to the DMM.
3.
Proceed to step 6 in the
TCP0020 &
TCP2020 Test
section. (See page 67,
20 A Current Probes Instruction Manual
65