Probing Principles
Maximum Current Limits
Current probes have three maximum current ratings: pulsed, continuous, and Ampere-second product. Exceeding any of these
ratings can saturate the probe core, which magnetizes the core and causes measurement errors. Refer to the speci
fi
cations for the
maximum current ratings of the probe. (See Table 2 on page 54.)
Maximum Pulsed Current (I
maxP
) is the
maximum peak value of pulsed current the
probe can accurately measure, regardless
of how short (within bandwidth limitations)
the pulse duration is.
Maximum Continuous Current (I
maxC
) is the
maximum current that can be continuously
measured at DC or at a speci
fi
ed AC
frequency. The maximum continuous
current value is derated with frequency;
as the frequency increases, the maximum
continuous current rating decreases.
20 A Current Probes Instruction Manual
43
Содержание TCP0020
Страница 1: ...x TCP0020 TCP2020 50 MHz 20 Amp AC DC Current Probes ZZZ Instruction Manual P071300200 071 3002 00...
Страница 2: ......
Страница 6: ......
Страница 10: ...Table of Contents iv 20 A Current Probes Instruction Manual...
Страница 22: ...Compliance Information xvi 20 A Current Probes Instruction Manual...
Страница 80: ...Specifications Figure 2 Typical insertion impedance versus frequency 56 20 A Current Probes Instruction Manual...
Страница 81: ...Specifications Figure 3 Maximum peak pulse versus pulse width 20 A Current Probes Instruction Manual 57...
Страница 83: ...Specifications Figure 4 Mechanical dimensions 20 A Current Probes Instruction Manual 59...