Performance Verification
40
P6209 4 GHz Probe Instruction Manual
NOTE
. Do not touch the probe tip adapter when making calibration
measurements. Measurement accuracy is degraded when the probe
tip adapter is handled.
8.
Use the oscilloscope measurement capability to calculate and
display rise time. Rise time is determined from the 10% and 90%
amplitude points on the waveform. Record the rise time as
t
s+p
.
9.
Calculate the probe only rise time using the following formula:
t
p
=
t
(
s
+
p
)
2
−
t
s
2
Ꭹ
10.
Check that the calculated rise time meets the probe specification.
The rise time (
t
p
) must be
≤
120 ps.
11.
Record the results on the test record.