Performance Verification
P6209 4 GHz Probe Instruction Manual
35
Output Zero
Use this procedure to verify the probe output zero.
1.
Connect the test equipment as shown in Figure 12.
2.
Ground the probe tip by connecting the two KlipChips together.
Digital
multimeter
BNC-to-dual
banana adapter
50
Ω
Precision
termination
CSA/TDS8000
SMA F-to-BNC M
adapter
Y-lead
adapter
KlipChip
adapter
P6209 Probe
Figure 12: P6209 output zero setup
3.
Set the oscilloscope trigger source to internal clock.
4.
Set the oscilloscope channel vertical offset to 0.00 V.
5.
Set the multimeter to read DC volts.
6.
Record the multimeter reading on the test record.
The measured DC level should be 0.00 V
±
10 mV for
the CSA/TDS8000.
Содержание P6209
Страница 5: ......
Страница 9: ...Table of Contents iv P6209 4 GHz Probe Instruction Manual...
Страница 21: ...Product Description 8 P6209 4 GHz Probe Instruction Manual...
Страница 29: ...Operating Basics 16 P6209 4 GHz Probe Instruction Manual...
Страница 33: ...Helpful Hints 20 P6209 4 GHz Probe Instruction Manual...
Страница 41: ......
Страница 45: ...Theory of Operation 32 P6209 4 GHz Probe Instruction Manual...
Страница 55: ...Performance Verification 42 P6209 4 GHz Probe Instruction Manual...
Страница 61: ...Adjustments 48 P6209 4 GHz Probe Instruction Manual...
Страница 63: ...Maintenance 50 P6209 4 GHz Probe Instruction Manual...
Страница 67: ...Troubleshooting 54 P6209 4 GHz Probe Instruction Manual...
Страница 75: ...Replaceable Parts P6209 4 GHz 5X Active Probe 62...