P6209 4 GHz Probe Instruction Manual
33
Performance Verification
Use the following procedures to verify the warranted specifications
of the P6209 probe. Before beginning these procedures, refer to
page 41 and photocopy the test record and use it to record the
performance test results. The recommended calibration interval is
one year.
These procedures test the following specifications:
H
Output offset voltage
H
DC attenuation accuracy
H
Rise time
Equipment Required
Refer to Table 5 for a list of the equipment required to verify the
performance of your probe.
NOTE
. These procedures use a CSA/TDS8000 as the host instrument.
If you are using a TDS 820, the procedures will vary slightly.
Table 5: Equipment required for performance verification and adjustment
Item description
Performance
requirement
Recommended example
Oscilloscope
TEKPROBE SMA interface
Tektronix CSA/TDS8000
with a 80E00 Series
Sampling Module, or
TDS 820
Calibration Step Generator
067-1338-XX
DC Power Supply
1
1 VDC at 1 mA
Tektronix PS280
DMM
0.5% accuracy at 2 VDC
Tektronix TX3
Содержание P6209
Страница 5: ......
Страница 9: ...Table of Contents iv P6209 4 GHz Probe Instruction Manual...
Страница 21: ...Product Description 8 P6209 4 GHz Probe Instruction Manual...
Страница 29: ...Operating Basics 16 P6209 4 GHz Probe Instruction Manual...
Страница 33: ...Helpful Hints 20 P6209 4 GHz Probe Instruction Manual...
Страница 41: ......
Страница 45: ...Theory of Operation 32 P6209 4 GHz Probe Instruction Manual...
Страница 55: ...Performance Verification 42 P6209 4 GHz Probe Instruction Manual...
Страница 61: ...Adjustments 48 P6209 4 GHz Probe Instruction Manual...
Страница 63: ...Maintenance 50 P6209 4 GHz Probe Instruction Manual...
Страница 67: ...Troubleshooting 54 P6209 4 GHz Probe Instruction Manual...
Страница 75: ...Replaceable Parts P6209 4 GHz 5X Active Probe 62...