Functional Check
12
P6209 4 GHz Probe Instruction Manual
4.
Set the offset to
1V
, the vertical scale to
500 mV/div
, and the
horizontal scale to
2
s/div
.
5.
The observed waveform should be a 200 kHz square wave with
an amplitude of approximately 2 V (the clock output is
unterminated), and have an offset of approximately 1 V.
If the probe does not pass this functional check, go to
Troubleshoot-
ing