Introduction
The information in this section deals with routine type maintenance that can be
performed by the operator.
Line fuse replacement
WARNING
Disconnect the line cord at the rear panel, and remove all test leads
connected to the instrument (front and rear) before replacing the line
fuse.
The power line fuse is accessible from the rear panel, just above the AC power receptacle (see
Figure 3-1).
Perform the following steps to replace the line fuse:
1. Carefully grasp and squeeze together the locking tabs that secure the fuse carrier to the
fuse holder.
2. Pull out the fuse carrier, and replace the fuse with the type specified in Table 3-1.
CAUTION
To prevent instrument damage, use only the fuse type specified in Table 3-1.
3. Re-install the fuse carrier.
NOTE
If the power line fuse continues to blow, a circuit malfunction exists and must be
corrected. Refer to the troubleshooting section of this manual for additional
information.
Table 3-1
Power line fuse
Line voltage
Rating
Keithley part
no.
100-240V
250V, 2.5A, slow blow 5
×
20mm
FU-106-2.5
3-2
Routine Maintenance
2410-902-01.book Page 2 Monday, November 7, 2005 2:49 PM
Содержание Keithley SourceMeter 2410
Страница 11: ...1 Performance Verification 2410 902 01 book Page 1 Monday November 7 2005 2 49 PM ...
Страница 29: ...2 Calibration 2410 902 01 book Page 1 Monday November 7 2005 2 49 PM ...
Страница 53: ...3 Routine Maintenance 2410 902 01 book Page 1 Monday November 7 2005 2 49 PM ...
Страница 56: ...4 Troubleshooting 2410 902 01 book Page 1 Monday November 7 2005 2 49 PM ...
Страница 74: ...5 Disassembly 2410 902 01 book Page 1 Monday November 7 2005 2 49 PM ...
Страница 84: ...6 Replaceable Parts 2410 902 01 book Page 1 Monday November 7 2005 2 49 PM ...
Страница 100: ...A Specifications 2410 902 01 book Page 1 Monday November 7 2005 2 49 PM ...
Страница 102: ...B Command Reference 2410 902 01 book Page 1 Monday November 7 2005 2 49 PM ...
Страница 116: ...C Calibration Programs 2410 902 01 book Page 1 Monday November 7 2005 2 49 PM ...