Table 2-4
Front panel current calibration
Source
range*
Source
current
Multimeter current
reading**
1µA
10µA
100µA
1mA
20mA
100mA
1A
+1.0000µA
+0.0000µA
-1.0000µA
-0.0000µA
+10.000µA
+00.000µA
-10.000µA
-00.000µA
+100.00µA
+000.00µA
-100.00µA
-000.00µA
+1.0000mA
+0.0000mA
-1.0000mA
-0.0000mA
+20.000mA
+00.000mA
-20.000mA
-00.000mA
+100.00mA
+000.00mA
-100.00mA
-000.00mA
+1.0000A
+0.0000A
-1.0000A
-0.0000A
____________ µA
____________ µA
____________ µA
____________ µA
____________ µA
____________ µA
____________ µA
____________ µA
____________ µA
____________ µA
____________ µA
____________ µA
___________ mA
___________ mA
___________ mA
___________ mA
___________ mA
___________ mA
___________ mA
___________ mA
___________ mA
___________ mA
___________ mA
___________ mA
_____________ A
_____________ A
_____________ A
_____________ A
*
*Use EDIT and RANGE keys to select source range.
**Multimeter reading used in corresponding calibration
step. See procedure.
2-14
Calibration
2410-902-01.book Page 14 Monday, November 7, 2005 2:49 PM
Содержание Keithley SourceMeter 2410
Страница 11: ...1 Performance Verification 2410 902 01 book Page 1 Monday November 7 2005 2 49 PM ...
Страница 29: ...2 Calibration 2410 902 01 book Page 1 Monday November 7 2005 2 49 PM ...
Страница 53: ...3 Routine Maintenance 2410 902 01 book Page 1 Monday November 7 2005 2 49 PM ...
Страница 56: ...4 Troubleshooting 2410 902 01 book Page 1 Monday November 7 2005 2 49 PM ...
Страница 74: ...5 Disassembly 2410 902 01 book Page 1 Monday November 7 2005 2 49 PM ...
Страница 84: ...6 Replaceable Parts 2410 902 01 book Page 1 Monday November 7 2005 2 49 PM ...
Страница 100: ...A Specifications 2410 902 01 book Page 1 Monday November 7 2005 2 49 PM ...
Страница 102: ...B Command Reference 2410 902 01 book Page 1 Monday November 7 2005 2 49 PM ...
Страница 116: ...C Calibration Programs 2410 902 01 book Page 1 Monday November 7 2005 2 49 PM ...