1-10
Performance Verification
Taking the SourceMeter out of compliance
Verification measurements should not be made when the SourceMeter is in compliance. For
purposes of the verification tests, the SourceMeter can be taken out of compliance by going into
the edit mode and increasing the compliance limit.
NOTE
Do not take the unit out of compliance by decreasing the source value or changing
the range. Always use the recommended range and source settings when
performing the verification tests.