![Tektronix Keithley SourceMeter 2410 Скачать руководство пользователя страница 27](http://html1.mh-extra.com/html/tektronix/keithley-sourcemeter-2410/keithley-sourcemeter-2410_service-manual_1078031027.webp)
Performance Verification 1-17
Resistance measurement accuracy
Follow the steps below to verify that Model 2410 resistance measurement accuracy is within
specified limits. This procedure involves applying accurate resistances from a resistance
calibrator and then verifying that Model 2410 resistance measurements are within required limits.
1.
With the power off, connect the resistance calibrator to the Model 2410 INPUT/
OUTPUT and 4-WIRE SENSE jacks as shown in Figure 1-3. Be sure to use the four-
wire
connections as shown.
2.
Select the resistance calibrator external sense mode.
3.
Configure the Model 2410 ohms function for the 4-wire sense mode as follows:
• Press CONFIG then MEAS
Ω
. The instrument will display the following:
CONFIG
OHMS
SOURCE SENSE-MODE GUARD
• Select SENSE-MODE, and then press ENTER. The following will be displayed:
SENSE-MODE
2-
WIRE 4-WIRE
• Select 4-WIRE, and then press ENTER.
• Press EXIT to return to normal display.
4.
Press MEAS
Ω
to select the ohms measurement function, and make sure the source
out-put is turned on.
5.
Verify ohms measurement accuracy for each of the resistance values listed in Table 1-6.
For each measurement:
• Set the resistance calibrator output to the nominal resistance or closest available
value.
NOTE
It may not be possible to set the resistance calibrator to the specified value. Use
the closest possible setting, and modify reading limits accordingly.
• Select the appropriate ohms measurement range with the RANGE keys.
• Verify that the Model 2410 resistance reading is within the limits given in the table.
6. Repeat the entire procedure using the rear panel INPUT/OUTPUT and 4-WIRE SENSE
jacks. Be sure to select the rear panel jacks with the front panel TERMINALS key.
2410-902-01.book Page 17 Monday, November 7, 2005 2:49 PM
Содержание Keithley SourceMeter 2410
Страница 11: ...1 Performance Verification 2410 902 01 book Page 1 Monday November 7 2005 2 49 PM ...
Страница 29: ...2 Calibration 2410 902 01 book Page 1 Monday November 7 2005 2 49 PM ...
Страница 53: ...3 Routine Maintenance 2410 902 01 book Page 1 Monday November 7 2005 2 49 PM ...
Страница 56: ...4 Troubleshooting 2410 902 01 book Page 1 Monday November 7 2005 2 49 PM ...
Страница 74: ...5 Disassembly 2410 902 01 book Page 1 Monday November 7 2005 2 49 PM ...
Страница 84: ...6 Replaceable Parts 2410 902 01 book Page 1 Monday November 7 2005 2 49 PM ...
Страница 100: ...A Specifications 2410 902 01 book Page 1 Monday November 7 2005 2 49 PM ...
Страница 102: ...B Command Reference 2410 902 01 book Page 1 Monday November 7 2005 2 49 PM ...
Страница 116: ...C Calibration Programs 2410 902 01 book Page 1 Monday November 7 2005 2 49 PM ...