:DATA?
(:CALibration:PROTected:SENSe:DATA?)
(:CALibration:PROTected:SOURce:DATA?)
Purpose
To request the calibration constants for the active range.
Format
:cal:prot:sens:data?
:cal:prot:sour:data?
Response
Four comma-separated ASCII floating-point constants.
Description
The :CAL:PROT:SENS:DATA? and :CAL:PROT:SOUR:DATA? queries
request the calibration constants for the active range of the sense and source
functions respectively. The four returned constants are in ASCII floating-
point format delimited by commas.
Note
To request the appropriate constants, (1) select the source or sense function,
(2) choose the desired range, and (3) send the desired :DATA? query.
Example
:SENS:FUNC 'VOLT:DC'
Select DC sense function.
:SENS:VOLT:DC:RANGE 0.2
Choose 200mV range.
:CAL:PROT:SENS:DATA?
Request cal constants.
B-10
Command Reference
2410-902-01.book Page 10 Monday, November 7, 2005 2:49 PM
Содержание Keithley SourceMeter 2410
Страница 11: ...1 Performance Verification 2410 902 01 book Page 1 Monday November 7 2005 2 49 PM ...
Страница 29: ...2 Calibration 2410 902 01 book Page 1 Monday November 7 2005 2 49 PM ...
Страница 53: ...3 Routine Maintenance 2410 902 01 book Page 1 Monday November 7 2005 2 49 PM ...
Страница 56: ...4 Troubleshooting 2410 902 01 book Page 1 Monday November 7 2005 2 49 PM ...
Страница 74: ...5 Disassembly 2410 902 01 book Page 1 Monday November 7 2005 2 49 PM ...
Страница 84: ...6 Replaceable Parts 2410 902 01 book Page 1 Monday November 7 2005 2 49 PM ...
Страница 100: ...A Specifications 2410 902 01 book Page 1 Monday November 7 2005 2 49 PM ...
Страница 102: ...B Command Reference 2410 902 01 book Page 1 Monday November 7 2005 2 49 PM ...
Страница 116: ...C Calibration Programs 2410 902 01 book Page 1 Monday November 7 2005 2 49 PM ...